| Test strip |
GE100 |
| Coding |
Auto-coding |
| Enzyme / Measurement technology |
GOD / Electrochemical sensor |
| Blood sample source |
Capillary whole blood (finger, palm and forearm) |
| Minimum sample volume |
0.75 µL |
| HCT range |
30 – 57% |
| Measuring range |
0.6 – 33.3 mmol/L (10 – 600 mg/dL) |
| Meter dimensions (L x W x H) |
98.0 x 46.0 x 17.5 mm (3.85 x 1.81 x 0.69 inch) |
| Display area (L x W) |
53.0 x 32.0 mm (2.09 x 1.26 inch) |
| Weight (with battery) |
53.0 ± 5 g |
| Power supply |
1 × CR2032 battery |
| Memory capacity |
500 blood glucose test results with date and time |
| Operating temperature |
10 – 40°C (50 – 104°F) |
| Operating relative humidity |
10 – 90% |
| Meter storage conditions |
-10 – 60°C (14 – 140°F) |
| Test strip storage conditions |
4 – 30°C (39 – 86°F), < 90% relative humidity |
| Accuracy |
95% of the measured glucose values shall fall within either
Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15% |
| Precision |
Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5% |
Fully compliant with ISO 15197:2013. (For other received credentials, please contact us.)