Test strip |
GT333 |
Coding |
Auto-coding |
Enzyme / Measurement technology |
GOD / Electrochemical sensor |
Blood sample source |
Capillary whole blood (finger, palm and forearm) |
Minimum sample volume |
0.75 µL |
HCT range |
20 – 60% |
Measuring range |
0.6 – 33.3 mmol/L (10 – 600 mg/dL) |
Meter dimensions (L x W x H) |
85.0 x 57.0 x 22.5 mm |
Display area (L x W) |
39.0 x 39.5 mm |
Weight (with battery) |
81.0 g |
Power supply |
2 × AAA batteries |
Memory capacity |
500 blood glucose test results with date and time |
Operating temperature |
10 – 40°C (50 – 104°F) |
Operating relative humidity |
10 – 90% |
Meter storage conditions |
-10 – 60°C (14 – 140°F) |
Test strip storage conditions |
4 – 30°C (39 – 86°F), <90% relative humidity |
Bluetooth version |
4.0 |
Accuracy |
95% of the measured glucose values shall fall within either
Glucose concentrations < 5.5 mmol/L (100 mg/dL), difference within ± 0.83 mmol/L (15 mg/dL)
Glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), bias within ± 15% |
Precision |
Glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)
Glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5% |
Fully compliant with ISO 15197:2013. (For other received credentials, please contact us.)