| Test strip |
Max Strip |
| Coding |
Auto-coding |
| Enzyme / Measurement technology |
GDH / Electrochemical sensor |
| Blood sample source |
Capillary, venous, arterial and neonatal whole blood |
| Minimum sample volume |
0.75 µL |
| HCT range |
10 – 70% |
| Measuring range |
0.6 – 33.3 mmol/L (10 – 600 mg/dL) |
| Meter dimensions (L x W x H) |
82.0 x 50.0 x 15.5 mm (3.23 x 1.97 x 0.61 inch) |
| Display area (L x W) |
40.7 x 40.2 mm (1.60 x 1.58 inch) |
| Weight (with battery) |
59.0 ± 5 g |
| Power supply |
1 × CR2032 battery for Max | 2 × CR2032 batteries for Max Plus |
| Memory capacity |
500 test results with date and time |
| Operating temperature |
6 – 44°C (43 – 111°F) |
| Operating relative humidity |
10 – 90% |
| Meter storage conditions |
-10 – 60°C (14 – 140°F) |
| Test strip storage conditions |
At 4 – 30°C (39 – 86°F), and below 90% relative humidity |
| Bluetooth version (For Max Plus) |
4.0 |
| Accuracy |
For glucose concentrations < 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements.
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 15% of reference measurements. |
| Precision |
For glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5% |
Fully compliant with ISO 15197:2013. (For other received credentials, please contact us.)