Test strip |
GS550 |
Coding |
Auto-coding |
Enzyme / Measurement technology |
GOD / Electrochemical sensor |
Blood sample source |
Capillary whole blood |
Minimum sample volume |
0.75 µL |
HCT range |
30 – 57% |
Measuring range |
0.6 – 33.3 mmol/L (10 – 600 mg/dL) |
Meter dimensions (L x W x H) |
90.6 x 46.0 x 16.5 mm (3.56 x 1.81 x 0.64 inch) |
Display area (L x W) |
46.6 x 33 mm (1.83 x 1.32 inch) |
Weight (with battery) |
53.0 ± 5 g |
Power supply |
2 × CR2032 batteries |
Memory capacity |
500 test results with date and time |
Operating temperature |
10 – 40°C (50 – 104°F) |
Operating relative humidity |
10 – 90% |
Meter storage conditions |
-10 – 60°C (14 – 140°F) |
Test strip storage conditions |
At 4 – 30°C (39 – 86°F), and below 90% relative humidity |
Accuracy |
For glucose concentrations < 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements.
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 15% of reference measurements. |
Precision |
For glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5% |
Fully compliant with ISO 15197:2013. (For other received credentials, please contact us.)